Check the possible causes and find out the solution!
Possible Causes
What You Can Do?
►
Improper standard dilution
Use appropriate diluent as blank. Make sure that the dilution is performed as according to protocol
►
Standard improperly reconstituted
Briefly spin standard vial before opening. Make sure that there is no undissolved material after reconstituting
►
Standard degraded
Store standards as according to protocol
►
Curve doesn’t fit the scale
Try plotting log-log or 5 parameter logistic curve fit
►
Pipetting error
Calibrate pipettes to make sure that the correct volume is dispensed
►
Incomplete washing
Increase washing cycles
-
Check the possible causes and find out the solution!
Possible Causes
What You Can Do?
►
Improper washing
Make sure that the washing is done as according to protocol
►
Poor mixing of samples
Mix samples gently and evenly
►
Dirty plate
Make sure that the bottom of plate is clean
►
Reagents too old
Make sure that the reagents are not expired. Use freshly prepared reagents
►
Bubbles in wells
Make sure that there is no bubble in wells before reading
►
Inconsistent pipetting
Calibrate pipettes to make sure that the correct volume is dispensed
►
Edge effects
Make sure that the plate and reagents are equilibrated to room temperature before starting assay
-
Check the possible causes and find out the solution!
Possible Causes
What You Can Do?
►
Too much antibodies was used
Reduce the concentration of primary or secondary antibodies
►
Antibodies bind nonspecifically
Use blocking buffer or choose another affinity-purified antibody
►
Too much substrate reagent used
Use substrate with higher dilution
►
Insufficient washing
Increase washing cycles
►
Wrong concentration of blocking reagent
Check the recommended concentration of blocking buffer
►
Reaction not stopped
Stop reactions with STOP buffer before reading
►
Plate left too long before reading
Take measurements shortly after addition of substrate and STOP buffer
►
Insufficient Tween in wash buffer
Use PBS+0.05% Tween as wash buffer
►
Incubation temperature too high
Optimize incubation temperature for each experiment
►
Plate stacking during incubation lead to uneven temperature throughout the plate
Avoid stacking plates together during incubation
►
Pipetting error
Calibrate pipettes to make sure that the correct volume is dispensed
►
Reagents not mixed properly
Make sure that all reagents are mixed properly and equilibrated to room temperature before assay
►
Salt concentration of incubation and wash buffer
Increase salt concentration to reduce nonspecific interaction
►
Substrate incubation carried out in light
Perform substrate incubation in dark
►
Dirty plate
Make sure that the bottom of plate is clean
-
Check the possible causes and find out the solution!
Possible Causes
What You Can Do?
►
Insufficient coating
Use more antigens or antibodies for coating
►
Substrate reagents have expired or prepared at a wrong pH
Use fresh substrate reagents
-
Check the possible causes and find out the solution!
Possible Causes
What You Can Do?
►
Assay set up incorrectly
Make sure that the instructions in the protocol is followed carefully
►
Incorrect secondary antibody used
Check if the correct secondary antibody is used
►
Insufficient antibodies used
Increase concentration of primary or secondary antibody
►
Substrate reagents not fresh
Use fresh substrate reagents
►
Wrong settings of plate reader
Check the settings (wavelength, filters, gain etc) of plate reader
►
Insufficient incubation
Follow the incubation time as indicated in the protocol booklet
►
Sample concentration falls below detection limits of kit
Decrease dilution factor or concentrate samples
►
Plate washing too vigorous
Check the setting of plate washer. Pipette wash buffer into wells gently
►
Wells dried out
Cover plate with adhesive film or incubate in humidified chamber throughout experiment
►
Enzyme inhibitor present in buffers or reagents
Inhibitors such as Sodium Azide can affect enzyme and assay performance. Ensure that there is no enzyme inhibitor in any buffers